    Index2 Fluke Series Pulse Oximeter Tester

    June 2026 1 min read

    مواصفات جهاز اختبار مقياس التأكسج Index2 Fluke Series والميزات ومنحنيات R للشركات المصنعة مثل Masimo و Nonin و Philips.

    The Index 2 is a lightweight, portable tool that includes preloaded manufacturers’ R-curves and the ability to define other “makes" for testing pulse oximeters.Index2 Fluke Series Pulse Oximeter Tester Motion presets, player mode, transmission level control (TLC), and computer commands boost testing ability. The Index 2 can also be configured to include an optional electrical simulation feature with probe test. Optical and electrical simulations allow technicians to isolate problems quickly. The probe test identifies defective probes with quantitative test results.   Features Portable Tenpreloaded manufacturers’ R-curves User-definable “makes for most other manufacturers R-curves for Masimo, Nonin and Philips Medical Systems (formerly Agilent / H-P) oximeters Six downloadable R-curve spaces available Simultaneous simulation of motion and arterial-oxygen levels Arterial wave-amplitude scale, calibrated in units of perfusion Tap/shiver motion simulations to explore the impact of motion RS-232 port for computer control Physiological finger   Electrical simulations with probe testing  
    Specifications
    O2
    Range: 35 % to 100 %
    Resolution: 1 %
    Accuracy: 100 % to 75 %: ± 1 % ± accuracy of the device under test; 74 % to 50 %: ± 3 % ± the accuracy of the device under test; < 50 % unspecified
    Repeatability: Repeatability: ± 1 standard deviation
    Rate
    Range: 30 BPM - 250 BPM
    Resolution: 1 BPM
    Accuracy: 1 % ± 1 BPM
    Pulse Amplitude
    Range: 0 to 100 % of nominal pulse amplitude
    Resolution: 1 %
    Probe test
    Continuity/Resistance: Pulse amplitude is 20 % of maximum pass-through brightness
    Test Matrix: Measures all combinations of possible interconnections in an XX point matrix
    Range: 250 O to 150 kO
    Accuracy: ± 5 % of reading
    LED/Detector voltage test
    Test Format: Measures the voltage drop across Red LED, infrared LED, and the photo detector when the internally generated test signal is applied
    Test Signal: Constant current source @ 1 mA
    Open Circuit: 2.5 V max
    Measurement/Display Range: 0 V to 4 V
    Accuracy: ± 5 % of reading, 0.4 V to 4 V
    Dynamic Test
    Test Format: Photodetector/diode response to both the red and infrared light generated by the probe when pulsed by an internal test signal
    Test Signal: Pulsed between the two LEDs; constant current level @ 1 mA
    Test Results: Nominal range of 0 to 2000
    Checksum
    Sum of all locations in the program chip; for service use only
    General Information
    Display: 2-line x 24-character super twist LCD
    Battery Life: At least 4 hours of continuous use
    Battery Charger: Lead acid battery charger only Input:100 V to 250 V, 50-60 Hz, 0.3 A Output:+12 V dc, 0.5 A
    Operating temperature: 15 oC to 35 oC (59 oF to 95 oF)
    Storage temperature: 0 oC to 50 oC (32 oF to 122 oF)
    Humidity 10% to 90% non-condensing
    Dimensions (WxDxH): 26.4 cm x 25.4 cm x 10.2 cm (10.4 in x 10 in x 4 in)
    Weight: 2.38 kg (5 lb)
    Share:

    More from News & Events

    View all articles
    Education
    Education Jun 2026

    Waterproof or water resistant? What’s behind an IP rating?

    Waterproof or water resistant

    Read more 1 min
    Education
    Education Jun 2026

    Reading ECGs (Electrocardiogram)in Dogs

    Reading ECGs (Electrocardiogram) in Dogs  

    Read more 1 min
    Education
    Education Jun 2026

    10 pravil učinkovite prve pomoči

    10 pravil učinkovite prve pomoči

    Read more 1 min