    How to calibrate Capnoxi Capnography Monitor

    June 2026 1 min read

    CapnOxi  meditech Capnograph Calibrating the CO2 Measurement It is recommended to calibrate CO2 measurement when you start CO2 monitoring

    CapnOxi  meditech Capnograph

    Calibrating the CO2 Measurement
    It is recommended to calibrate CO2 measurement when you start CO2 monitoring. But you must carry out the CO2 calibration when the CO2 need zero calibration displays in the alarm area.
     
    Procedure:
    1. Press menu key on the keyboard to enter the Main Menu dialog.
    2. Select User Settings button to enter the User Settings dialog.
    3. Select Calibrate Zero button to calibrate CO2 measurement.
    4. About 25 seconds later, if the calibration failed, the failure message will be shown.
     
    Share:

    More from News & Events

    View all articles
    Product Library
    Product Library Aug 2026

    MD901F vs MD901F-Android Fetal Maternal Monitor Comparison Table

      ParameterMD901FMD901F-AndroidType12.1 inch12.1 inchStandard ConfigurationFHR, TOCO, FMFHR, TOCO, FM, Touch Screen, NetworkingOptionsFHR2, SpO₂ + NIBP, Fetus Waker, Wireless FHR/TOCO/FM, Touch…

    Read more 1 min
    Product Library
    Product Library Jun 2026

    iBreathe TOUCH SCREEN SPIROMETER med inbyggd skrivare

    iBreathe TOUCH SCREEN SPIROMETER med inbyggd skrivare Bärbar enhet för lungfunktionstestning Denna spirometer använder differentialtrycksteknik för att noggrant mäta över…

    Read more 1 min
    Product Library
    Product Library Jun 2026

    How AED pads are manufactured?

    How AED pads are manufactured The production of AED pads is a meticulous process involving several stages to ensure their…

    Read more 1 min