Automatic External Defibrillator Defi5

| Medical and technical requirements AED Defi5 Defibrillators |
| Presence of a battery charge level indicator |
| Non-rechargeable LiMnO2 battery |
|
Battery capacity: 100 discharges at 200 J or 120 discharges at 150 J |
| Three-stage defibrillation process |
|
Has three defibrillation energies: 1. First energy level 150 J 2. Second energy level 150 J 3. Third energy level 200 J |
| Broad voice and visual prompts to protect against blocking of the biphasic energy output by the operator to prevent accidental defibrillation. |
| Presence of 22 voice prompts in Ukrainian |
| Presence of three battery charge level indicators |
| The device must operate only from the internal battery |
| Presence of audio instructions via a built-in speaker |
| recognizes ventricular fibrillation and other ventricular tachycardia and guides operators through the defibrillation process |
| Electrodes and ECG cables contain ferromagnetic materials |
| Set of two multifunctional electrodes |
|
Presence of monthly self-testing: charge and discharge test, data collection and data transmission check, battery test |
|
Defibrillator charging time 10 sec. to 150 J 15 sec. to 200 J |
| Output energy accuracy ±15% at any impedance from 25 to 175 Ohms |
| Patient impedance: 20 Ohms to 200 Ohms |
| RF emission: class B |
| CISPR: Not applicable |
| RF emission CISPR 11 Group 1 |
| Operating humidity: relative humidity from 30% to 95% (non-condensing) |
| Pressure: 102.4 kPa~57.2 kPa (768 mmHg~429 mmHg) |
|
Storage temperature (without battery): -20℃ … 55℃ |
| Weight no more than 2.0 kg |
| Shelf life: 5 years |
| The kit must include a case for the automated external defibrillator |
| Presence of two analysis results |
| Analysis must last 8~10 seconds |
| Weight 2.0 kg |
| Dimensions 303 x 216 x 89 mm |
| Synchronization function |