Display:
Type: high resolution;
Size: 8 inches;
Sweep speed: 25 mm/sec.
Information: Heart rate; Lead / Electrode, Alarm ON / OFF, SpO2; Automatic defibrillation and prompt functions; Alarm and limit selection; Charge delivered.
|
SpO2 Module (Optional):
Measurement range: 30~100%, ± 2% between 80% ~ 90%, others ± 5%;
Alarm range: selection between high or low limit;
Alarm accuracy: ±10% within the set values;
Alarm time accuracy: < 12 sec.
EtCO2 (Optional)
NIBP (Optional)
|
|
Defibrillator:
Waveform: biphasic;
Charge time: < 7 sec. with a new, fully charged battery;
Power: the display shows the selected and delivered power;
Prompt types: voice and visual;
Electrode impedance measurement range: 0 – 250 W.
|
ECG monitoring:
Patient connection: ECG cable with three or five inputs via electrodes;
Lead selection: monitor display, pads, I, II, III, AVR, AVL, AVF, V; ECG sensitivity: 0.25, 0.5, 1, 1.5, 2, 4 mm / mV displayed on the monitor;
Heart rate: 20 – 300 BPM;
Heart rate alarm: ON / OFF displayed on the monitor;
Smart alarms: audible signal/voice commands indicate shockable arrhythmia.
|
|
Recording device:
Paper: 50 mm thermal paper;
Speed: 12.5 mm/sec, 25 mm/sec, 50 mm/sec; option to select 6 sec delay;
Printing method: high-resolution thermal print head;
Printing modes: manual or automatic, with adjustable settings;
ON / OFF control: front panel and electrodes.
Automatic function: 9-second recording, triggered by alarm activation, defibrillation charge, or defibrillation shock.
|
Automatic defibrillation mode:
Automatic function: automatic analysis and charge x3 with pre-programmed automatic charge selection, on-screen commands, voice commands;
Automatic diagnosis of shockable arrhythmia: ventricular fibrillation with amplitude ≥ 200 mV, ventricular tachycardia with rate ≥ 400 bpm, QRS complex wave amplitude ≥ 140 ms;
Charge control: from the front panel; press the handle button;
Commands: visual and voice.
|
|
Battery:
Type: rechargeable Ni-MH, 12V;
Operating time: for a new, fully charged battery – 60 defibrillation shocks at maximum charge, or a minimum of three hours of continuous ECG monitoring. Additional parameters affect operating time in different modes.
|
Manual mode:
Charge selection at: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 J;
Synchronized mode: synchronization of the defibrillation impulse with the patient's R-wave; the display shows the message "SYNC".
|