EKG6012 6-Channel ECG
The internal memory of the 6-channel electrocardiograph EKG6012 allows storing up to 1000 recorded ECGs with their automatic interpretation and analysis of key parameters. ECG viewing on a large 800x400 px display. The built-in battery allows recording up to 150 ECGs.
Together with a PC, the 6-channel electrocardiograph EKG6012T forms a workstation for diagnosing a large patient flow.
6/12-channel electrocardiograph EKG6012T, package contents:
ECG device - 1 pc.
Instruction manual - 1 pc.
ECG cable - 1 pc.
ECG cable - 1 pc.
USB cable for connecting the electrocardiograph to a PC - 1 pc.
Paper - 1 pc.
Power cable - 1 pc.
Bulb electrode - 6 pcs.
Clamp electrode - 4 pcs.
Warranty card - 1 pc.
Optional package: carrying bag.
Technical specifications for the 6/12-channel electrocardiograph EKG6012T:
Protection type: Class II B, type CF, with defibrillation protection.
ECG cable: 10 standard electrodes, for acquiring 12 ECG leads.
General and defibrillation protection.
Patient leakage current: <10μA
Input impedance: ≥50MΩ
Sampling rate: 1000Hz
ADC resolution: 12 bit
Frequency response: 0.05Hz~150Hz(-3dB~0.4dB)
Time constant: >3.2s
CMRR (common-mode rejection ratio): >60dB >100 dB (with filter added)
Noise level: ≤15μV p-p
EMG filter: 25Hz/35Hz (-3dB)
AC filter: 50Hz/60Hz(≥20dB)
Thermal printer:
Thermal printing system (density 8 dots/mm)
Recording paper: 110mm(W)x20m(L) high-speed thermal printing paper.
Sensitivity: 2.5 mm/mV 5mm/mV 10mm/mV 20mm/mV 40mm/mV error: ±5%.
Standard sensitivity: 10mm/mV ±0.2 mm/mV
Automatic mode:
Automatic ECG recording in the current format, automatic lead switching, automatic measurement and analysis.
Additional mode with rhythmogram recording:
ECG recording in the current format plus rhythm recording, automatic measurement and analysis.
Manual recording mode:
ECG recording in the current format, manual lead switching.
Measured parameters: HR, PR interval, P wave, QRS complex, T wave, QT interval, QTc, P wave axis, QRS axis, T wave axis, R(V5), S(V1), R(V5)+ S(V1).
Paper speed:
In automatic mode: 25mm/s, 50mm/s error: ±5%
In rhythmogram recording mode: 25mm/s, 50mm/s error: ±5%
In manual mode: 5 mm/s, 6.25 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s error: ±5%
Display: color touchscreen display with a resolution of 800 x 480 pixels
Power supply: AC:220V, 50Hz (110V, 60Hz)
DC: 14.8V, 3700 mAh rechargeable lithium battery
Fuse specifications: 2 pcs. size 5x20mm AC; 1.6mA/250V(220V power supply)
Dimensions: 340mm x 320mm x 85mm
Weight: 3.2 kg
The device can record 150 recordings and continuously print for 90 min. after fully charging the battery.
Archive size of more than 1000 patient records.
Normal operating conditions:
a) Permissible temperature: +5°C ~+35°C
b) Permissible humidity:≤80%
d) Atmospheric pressure: 645 mmHg ~ 795 mmHg
Storage and transportation of the electrocardiograph:
a) Permissible temperature: -10°C ~+55°C
b) Permissible humidity: ≤95%
c) Atmospheric pressure: 375 mmHg ~ 795 mmHg
Synchronization with a computer:
Minimum system requirements for the computer:
Processor: Pentium PIII, Celeron 900 or higher.
RAM: min 128MB.
Motherboard: It is recommended to use motherboards with a system chipset manufactured by Intel.
Hard drive: from 10GB.
Graphics adapter (video card): with memory capacity from 16MB.
Printer: laser printer, print resolution not less than 600dpi.
Monitor: from 14 inches.
CD ROM: CD-ROM 24x or higher.
Operating system: Microsoft Windows 2000/ XP (Windows XP with 32-bit architecture is recommended).
Monitor resolution: from 1024*768 pixels.
Color: color depth from 24 bit.
Manufacturer Meditech Number of recorded channels 12 Simultaneous printing (channels) 6 Paper width 110 Paper type roll Display size 7" Display type TFT touchscreen, color On-screen display (channels) 12 Printing speed Auto recording and rhythm recording: 25; 50 mm/s Manual recording: 5; 6.25; 10; 12.5; 25; 50; 100 mm/s Keyboard (letters, numbers) no PC connection yes Software yes Interpretation yes Built-in battery yes Carrying bag option Dimensions 315x215x92