Monitor paziente multiparametrico
Descrizione prodotto: Monitor portatile con funzionamento a rete elettrica e a batterie ricaricabili
Read more
8-inch multi-parameter patient monitor with ECG, SpO₂, NIBP, and advanced analysis
8-Inch Screen Patient Monitor is a compact and reliable bedside monitoring solution designed for hospitals, clinics, and bedside applications. It provides comprehensive monitoring of essential vital signs through an intuitive touchscreen interface.
All Meditech customers and healthcare professionals can submit equipment service tickets through the Customer Service Portal.
To receive two additional months of extended warranty, please register your product here.To browse all support resources and contact information, please visit the Support Center.
Coming soon.
Discover the advanced capabilities that set our products apart
The 8-inch multi-parameter patient monitor is a high-precision clinical monitoring system equipped with a TFT touchscreen, designed for comprehensive monitoring of adult, pediatric and neonatal patients. With flexible waveform display options, advanced ECG analysis functions, and a rechargeable battery, the device provides reliable monitoring in hospitals, clinics and critical care units.
Detailed technical specifications and performance parameters
| Parameter | Specifications |
|---|---|
| Lead Configuration | 5 Leads (R, L, F, N, C / RA, LA, LL, RL, V) |
| Lead Selection | I, II, III, aVR, aVL, aVF, V |
| Waveform Display | 2 Channels |
| Lead Configuration | 3 Leads (R, L, F / RA, LA, LL) |
| Lead Selection | I, II, III |
| Waveform | 1 Channel |
| Gain | ×2.5, ×5, ×10, ×20 mm/mV, Automatic |
| Heart Rate Range | Adult: 15–300 bpm / Neo-Ped: 15–350 bpm |
| Heart Rate Accuracy | ±1% or ±1 bpm |
| Resolution | 1 bpm |
| Sensitivity | >200 μV |
| Differential Input Impedance | >5 MΩ |
| CMRR | Monitoring: >105 dB / Operating: >105 dB / Diagnostic: >85 dB |
| Electrode Offset Voltage | ±300 mV |
| Leakage Current | <10 μA |
| Baseline Recovery | <3 s after defibrillation |
| ECG signal range | ±8 mV (Vp-p) |
| Bandwidth | Surgery: 1–15 Hz / Monitoring: 0.5–35 Hz / Diagnostic: 0.05–100 Hz |
| Signal Calibration | 1 mV ±5% |
| ST Monitoring Range | -2.0 to +2.0 mV |
| Arrhythmia Detection | Asystole, VFib/VTAC, Couplet, Bigeminy, Trigeminy, R on T, VT >2, PVC, TACHY, BRADY, Missed Beats, PNP, PNC |
| Alarm Function | Available |
| Verification Function | Available |
| Parameter | Specifications |
|---|---|
| Measurement Method | Impedance between RF (RA-LL) |
| Differential Input Impedance | >2.5 MΩ |
| Impedance Measurement Range | 0.3–5.0 Ω |
| Impedance Baseline Range | 0–2.5 kΩ |
| Bandwidth | 0.3–2.5 Hz |
| Respiration Rate Range | Adult: 0–120 RPM / Neo-Ped: 0–150 RPM |
| Resolution | 1 RPM |
| Accuracy | ±2 RPM |
| Apnea Alarm | 10–40 s |
| Parameter | Specifications |
|---|---|
| Measurement method | Oscillometric |
| Modes | Manual / Automatic / STAT |
| Automatic measurement interval | 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 minutes |
| STAT mode duration | 5 minutes |
| Pulse rate range | 40–240 bpm |
| Alarm type | SYS, DIA, Mean |
| Patient mode | SYS | DIA | Mean |
|---|---|---|---|
| Adult | 40–270 mmHg | 10–215 mmHg | 20–235 mmHg |
| Pediatric | 40–200 mmHg | 10–150 mmHg | 20–165 mmHg |
| Neonatal | 40–135 mmHg | 10–100 mmHg | 20–110 mmHg |
| Parameter | Specifications |
|---|---|
| Accuracy | 1 mmHg |
| Accuracy | Mean deviation ≤±5 mmHg; standard deviation ≤±8 mmHg |
| Overpressure protection | Adult: 297±3 mmHg / Pediatric: 240±3 mmHg / Neonatal: 147±3 mmHg |
| Parameter | Specifications |
|---|---|
| Measurement range | 0–100% |
| Alarm range | 0–100% |
| Accuracy | 1% |
| Accuracy | 70–100%: ±2% / 0–69%: unspecified |
| Update interval | Approximately 1 second |
| Alarm delay | 10 s |
| Pulse rate range | 0–254 bpm |
| Pulse rate accuracy | 1 bpm |
| Pulse rate accuracy | ±2 bpm |
| Parameter | Specifications |
|---|---|
| Channel | Single channel |
| Measurement & Alarm Range | 0–50°C |
| Resolution | 0.1°C |
| Accuracy | ±0.1°C |
| Update Interval | Approximately 1 second |
| Average Time Constant | <10 s |
| Parameter | Specifications |
|---|---|
| Supported Parameters | ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
| Parameter | Range |
|---|---|
| ART | 0–300 mmHg |
| PA | -6–120 mmHg |
| CVP / RAP / LAP / ICP | -10–40 mmHg |
| P1 / P2 | -10–300 mmHg |
| Parameter | Specifications |
|---|---|
| Pressure Sensor Sensitivity | 5 μV/V/mmHg |
| Impedance | 300–3000 Ω |
| Accuracy | 1 mmHg |
| Accuracy | ±2% or ±1 mmHg |
| Response Time | Approx. 1 Second |
Compatible accessories to extend functionality
| Item | Quantity |
|---|---|
| ECG Cable (5 Leads) | 1 Piece |
| SpO₂ Cable (Adult) | 1 Piece |
| TEMP Probe (Adult) | 1 Piece |
| NIBP Cuff (Adult) | 1 Piece |
| NIBP Tube (150 cm) | 1 Piece |
| Grounding Cable | 1 Piece |
| Power Cable | 1 Piece |
| Cleaning Cloth | 1 Piece |